Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes, Intraoperative Subdural Electrodes
Electrode, Cortical. 510(k) K053363, from Ad-Tech Medical Instrument Corp (Stoughton, WI). Product code GYC, class 2.
As of Nov 18, 2026, K053363 is an FDA 510(k) held by Ad-Tech Medical Instrument Corp, substantially equivalent on May 30, 2006 after 179 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 30, 2006
- Review time: 179 days
- Recalls: 2
- Cited as a predicate by: 6 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe AD-TECH Subdural Electrodes (Dual-Sided Interhemispheric, Grid, Intraoperative, Strip, Wyler) are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals on the surface of the brain.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ad-Tech Subdural ElectrodeK970587 · Ad-Tech Medical Instrument Corp · May 9, 1997
- Ad-Tech Sub Dural ElectrodedsK923803 · Ad-Tech Medical Instrument Corp · Oct 23, 1992
- Subdural Strip ElectrodeK021144 · Nicolet Biomedical, Inc. · Jun 24, 2002
Cited as a Predicate
6 later 510(k) summaries name this device as a predicate- WISE Cortical Strip (WCS)K221123 · Wise S.R.L. · Nov 22, 2022
- Blackrock NeuroCoG Subdural Cortical Electrodes (Strips and Grids)K191346 · Blackrock Microsystems · Feb 14, 2020
- Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative Grid Electrode, Dual-Sided Interhemispheric…K191186 · Ad-Tech Medical Instrument Corporation · Jan 25, 2020
- AirRay Subdural Cortical ElectrodesK183437 · Cortec GmbH · Mar 12, 2019
- Neuromaster G1 MEE200K142624 · Nihon Kohden Corporation · Apr 24, 2015
- Pmt Subdural Cortical ElectrodesK082474 · Pmt Corp. · May 8, 2009
Recalls
Recalls FDA ties to K053363, newest first; each opens FDA's recall record| Jun 18, 2019 | Z-2406-2019Supplemental information provided with devices indicates that the subdural electrodes, depth electrodes, and anchor bolts are approved for use in the MR environment, for which they are not approved for such use. | Terminated |
| Dec 18, 2012 | Z-0892-2013The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with… | Terminated |
Other GYC Clearances
The newest 510(k)s for the same product code; their median review was 218 days; see every GYC device- Atlas Stim Headbox (NK) (31-0601-0077); Atlas Stim Headbox (Touchproof) (31-0601-0089); Atlas Stim Headbox (HD)…K250094 · Neuralynx, Inc. · Apr 14, 2025
- Layer 7-TK242618 · Precision Neuroscience, Corp. · Mar 30, 2025
- WISE Cortical Strip (WCS)K221123 · Wise S.R.L. · Nov 22, 2022
- Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid, Multi-Strip and Split Grid, Intraoperative…K211954 · Spes Medica Srl · Nov 10, 2022
- Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)K201931 · Dixi Medical · Oct 2, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Dec 2, 2005
- Review panel
- Neurology
- Regulation
- 21 CFR 882.1310
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files