Modification To: NxStage Pureflow-B Solution
Dialysate Concentrate For Hemodialysis (Liquid Or Powder). 510(k) K053286, from Nxstage Medical, Inc. (Lawrence, MA). Product code KPO, class 2.
As of Nov 20, 2026, K053286 is an FDA 510(k) held by Nxstage Medical, Inc., substantially equivalent on Dec 22, 2005 after 27 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 27 days
- Recalls: 14
- Cited as a predicate by: 1 device
NoStage PureFlow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.
- pureFLOW 402, pureFLOW 400, pureFLOW 406, pureFLOW 401, pureFLOW 502, pureFLOW 504K233159 · Fresenius Medical Care Renal Therapies Group, LLC · Dec 1, 2023
| Apr 28, 2025 | Z-1864-2025Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior… | Open |
| May 14, 2024 | Z-2160-2024The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than… | Open |
| Apr 11, 2019 | Z-1706-2019, Z-1707-2019Certain lots were mislabeled with incorrect product number. | Terminated |
| Oct 17, 2018 | Z-0505-2019, Z-0500-2019, Z-0504-2019, Z-0507-2019, Z-0503-2019, Z-0501-2019, Z-0506-2019, Z-0502-2019PureFlow B Solution smaller chamber of the two chamber bag can burst electrolyte fluid and cause injury if it comes in contact with the eyes or skin | Terminated |
| Nov 10, 2017 | Z-1132-2018Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk. The NxStage PureFlow B Solution is for use with systems that use one premixed dialysate… | Terminated |
| Jan 24, 2014 | Z-1073-2014Mislabeled product. | Terminated |
- Nipro Dry Acid Concentrate MixerK260799 · Nipro Technical Solutions, Incoporated · Aug 19, 2026
- NaturaLyte® Dry Bicarbonate Concentrate (08-4112-2)K253462 · Fresenius Medical Care Renal Therapies Group, LLC · Jan 29, 2026
- Bicarby Dialysate RFP-404 (RFP-404-W); Bicarby Dialysate RFP-403 (RFP-403-W); Bicarby Dialysate RFP-403 (RFP-403-G);…K252180 · Fresenius Medical Care Renal Therapies Group, LLC · Sep 9, 2025
- Bicarby Dialysate RFP-402 (RFP-402-G); Bicarby Dialysate RFP-400 (RFP-400-G); Bicarby Dialysate RFP-407 (RFP-407-G);…K243786 · Fresenius Medical Care Renal Therapies Group, LLC · Apr 4, 2025
- HemoCare Bicarbonate Concentrate Set (BCS)K240920 · Deka Research and Development · Jul 12, 2024
- pureFLOW 402 (F00012067); pureFLOW 406 (F00012068); pureFLOW 401 (F00012069); pureFLOW 400 (F00012070)K233950 · Fresenius Medical Care Renal Therapies Group, LLC · May 10, 2024
- NxStage PureFlow SolutionK233213 · Nxstage Medical, Inc. · Jan 18, 2024
- pureFLOW 402, pureFLOW 400, pureFLOW 406, pureFLOW 401, pureFLOW 502, pureFLOW 504K233159 · Fresenius Medical Care Renal Therapies Group, LLC · Dec 1, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Nov 25, 2005
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.5820
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files