Arca
Skip to content
Tools
Tools

Pvs 2000 Synchro2 Guidewire

Wire, Guide, Catheter. 510(k) K053268, from Boston Scientific - Precision Vascular (West Valley City, UT). Product code DQX, class 2.

As of Nov 19, 2026, K053268 is an FDA 510(k) held by Boston Scientific - Precision Vascular, substantially equivalent on Mar 13, 2006 after 110 days, with 13 recalls tied to it in FDA's recall database.

  • Review time: 110 days
  • Recalls: 13
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The PV 2000 Synchro®2 Guidewire series of products are intended for general intravascular use, including the neuro and peripheral vasculature. The device can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K053268, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 18, 2024Z-1941-2024Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer…Open
Jun 7, 2010Z-0872-2011, Z-0873-2011, Z-0875-2011, Z-0871-2011, Z-0869-2011, Z-0868-2011, Z-0870-2011, Z-0874-2011The Guidewire has the potential to flake off polytetrafluroethylene (PTFE) particulate, which has the potential to embolize and result in infarction when introduced into a patient's circulatory system.Terminated
Jun 3, 2010Z-2069-2010, Z-2070-2010, Z-2068-2010, Z-2071-2010Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the…Terminated

Other DQX clearances

The newest 510(k)s for the same product code; their median review was 151 days; see every DQX device

Record

Type
Special
Received
Nov 23, 2005
Review panel
Cardiovascular
Regulation
21 CFR 870.1330

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files