Tram Module
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K053121, from GE Healthcare (Needham, MA). Product code MHX, class 2.
As of Nov 21, 2026, K053121 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Apr 5, 2006 after 149 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 149 days
- Recalls: 2
The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another.
- Tram 2001 ModuleK011000 · General Electric Medical Systems Information Techn · Jun 29, 2001
| Dec 5, 2013 | Z-0808-2014Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection… | Terminated |
| Apr 3, 2013 | Z-2117-2013GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module. ECG… | Terminated |
- 12SL PrecisionK261307 · Ge Medical Systems Information Technologies, Inc. · Nov 3, 2026
- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) – VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Nov 7, 2005
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files