Arca
Skip to content

Stryker Spine Trio Spinal Fixation System

Orthosis, Spinal Pedicle Fixation. 510(k) K052971, from Stryker Spine (Allendale, NJ). Product code MNI, class 2.

As of Nov 17, 2026, K052971 is an FDA 510(k) held by Stryker Spine, substantially equivalent on Nov 16, 2005 after 23 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 16, 2005
  • Review time: 23 days
  • Recalls: 2
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Stryker Spine TRIO® Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K052971, newest first; each opens FDA's recall record
Jul 1, 2009Z-1886-2009The Trio 3.5MM Hex socket has been reported to break under high stresses.Terminated
Sep 5, 2006Z-0986-2009Reports were received of the ball ring ejecting from the connector's body when other connectors have already been locked down.Terminated

Record

Type
Special
Received
Oct 24, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files