RX Imola
Electrode, Ion Specific, Potassium. 510(k) K052914, from Randox Laboratories, Ltd. (Crumlin, GB). Product code CEM, class 2.
As of Nov 20, 2026, K052914 is an FDA 510(k) held by Randox Laboratories, Ltd., substantially equivalent on Mar 31, 2006 after 168 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 168 days
- Recalls: 4
- Cited as a predicate by: 2 devices
The RX Imola is a medium-sized desktop fully Automated Clinical Chemistry Analyzer complete with Ion Selective Electrode (ISE) Unit and dedicated analyzer software. An external PC operates the analyzer and results can be printed as required. The analyzer may be connected to a host computer, when required. The analyzer can be used to run tests such as magnesium in serum and plasma samples.
- ISE ElectrodesK230890 · Randox Laboratories, Ltd. · Sep 8, 2023
- RX Daytona Plus Chemistry Analyzer; RX Daytona Plus Aspartate Aminotransferase (Ast) ReagentK131554 · Randox Laboratories, Ltd. · Jan 9, 2014
| Sep 6, 2023 | Z-0200-2024RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature… | Open |
| Jun 24, 2019 | Z-2084-2019There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument. | Terminated |
| May 4, 2018 | Z-2059-2018Abnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results. | Terminated |
| Apr 14, 2016 | Z-0725-2017According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid… | Terminated |
- ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutionsK253491 · Roche Diagnostics · Feb 12, 2026
- ABL90 FLEX PLUS SystemK241037 · Radiometer Medicals Aps · Jan 14, 2025
- EasyStat 300K220396 · Medica Corporation · Sep 26, 2023
- ISE ElectrodesK230890 · Randox Laboratories, Ltd. · Sep 8, 2023
- Type
- Traditional
- Received
- Oct 14, 2005
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files