Syngo Imaging Version V20A
System, Image Processing, Radiological. 510(k) K052461, from Siemens AG Medical Solutions (Erlangen, DE). Product code LLZ, class 2.
As of Nov 18, 2026, K052461 is an FDA 510(k) held by Siemens AG Medical Solutions, substantially equivalent on Sep 22, 2005 after 14 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 14 days
- Recalls: 3
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFsyngo® Imaging is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital radiological images, including digital mammography images. It supports the physician in diagnosis and treatment planning. syngo® Imaging also supports storage and archiving of Structured DICOM Reports.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Sienet Cosmos (Version V15A)K042832 · Siemens AG Medical Solutions · Oct 28, 2004
- LeonardoK040970 · Siemens Medical Systems, Inc. · Jul 8, 2004
- Mammoreport Softcopy WorkstationK042868 · Siemens Medical Solutions USA, Inc. · Jan 6, 2005
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Chili PacsK070796 · Chili GmbH · Apr 6, 2007
- MediportK060210 · Maxant Technologies, Inc. · Feb 10, 2006
Recalls
Recalls FDA ties to K052461, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 3, 2014 | Z-1622-2014A complaint reported that the archiving method at one customer site was changed from implicit to explicit, and the cleanup procedure enabled by the explicit archivingmethod may have deleted data from archive itself. | Terminated |
| Aug 22, 2008 | Z-0102-2009Image may not be visible after merge. | Terminated |
| Sep 5, 2006 | Z-0070-2007Potential issue with annotation function. Annotations, shown as overlays, may change font size when the image is saved. Text may appear shifted, or lines appear thicker, and their position may not be as intended. | Terminated |
Other LLZ clearances
The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Sep 8, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files