Arca
Skip to content
Tools
Tools

Dorsal Nail Plate, Anatomical

Plate, Fixation, Bone. 510(k) K052248, from Hand Innovations, LLC (Miami Lakes, FL). Product code HRS, class 2.

As of Nov 18, 2026, K052248 is an FDA 510(k) held by Hand Innovations, LLC, substantially equivalent on Sep 14, 2005 after 27 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 27 days
  • Recalls: 5

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Dorsal Nail Plate™, Anatomical is intended for the fixation of fractures and osteotomies involving the distal Radius.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K052248, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 23, 2014Z-2212-2014The firm received complaints indicating that the drill guide becomes stuck in the jig. The drill guide can be placed in the jig about halfway before it becomes stuck.Terminated
Dec 23, 2008Z-0984-2009, Z-0982-2009, Z-0983-2009, Z-0981-2009The DNP nail plate may not attach to the alignment guide (jig) instrument, making it difficult to insert the nail plate into position or to drill needed holes accurately.Terminated

Other HRS clearances

The newest 510(k)s for the same product code; their median review was 73 days; see every HRS device

Record

Type
Special
Received
Aug 18, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files