Solanas Posterior Stabilization System
Appliance, Fixation, Spinal Interlaminal. 510(k) K052201, from Alphatec Spine, Inc. (Carlsbad, CA). Product code KWP, class 2.
As of Nov 17, 2026, K052201 is an FDA 510(k) held by Alphatec Spine, Inc., substantially equivalent on Nov 10, 2005 after 90 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 10, 2005
- Review time: 90 days
- Recalls: 3
- Cited as a predicate by: 7 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIt is intended that this device, in any system configuration, be removed after development of solid fusion mass. Hook components are indicated for use at C1-C7. Polyaxial screws are limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. These screws are not intended for placement in the cervical spine. 1.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Summit Occipito-Cervico-Thoracic (Oct) Spinal SystemK002733 · Depuyacromed · Dec 15, 2000
Cited as a Predicate
7 later 510(k) summaries name this device as a predicate- Blackbird Spinal SystemK133214 · Choicespine, LP · Mar 7, 2014
- Solanas Posterior Stabilization SystemK131119 · Alphatec Spine, Inc. · May 21, 2013
- Reliance Posterior Cervical- Thoracic SystemK122292 · Reliance Medical Systems, LLC · Oct 24, 2012
- Spinal Elements Posterior Cervical/Thoracic Spinal SystemK120467 · Spinal Elements, Inc. · Jul 30, 2012
- Sure Lok Mini Posterior Cervical/Upper Thoracic SystemK112025 · Spinal USA · Sep 15, 2011
- Lanx Posterior Cervicothoracic Spinal Fixation System (Pcfs)K100191 · Lanx, Inc. · Apr 15, 2010
- Modification To Solanas Posterior Stabilization SystemK071380 · Alphatec Spine, Inc. · Aug 20, 2007
Recalls
Recalls FDA ties to K052201, newest first; each opens FDA's recall record| Nov 1, 2010 | Z-0932-2012These lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips. | Terminated |
| Aug 5, 2009 | Z-0502-2010Alphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40… | Terminated |
| Aug 5, 2009 | Z-0503-2010Alphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 12, 2005
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files