Arca
Skip to content
Tools
Tools

Opti Lion Electrolyte Analyzer, Model GD7200

Electrode, Ion-Specific, Chloride. 510(k) K052027, from Osmetech, Inc. (Roswell, GA). Product code CGZ, class 2.

As of Nov 18, 2026, K052027 is an FDA 510(k) held by Osmetech, Inc., substantially equivalent on Sep 29, 2005 after 64 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 29, 2005
  • Review time: 64 days
  • Recalls: 4

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K052027, newest first; each opens FDA's recall record
Aug 9, 2013Z-0029-2014OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported…Terminated
Mar 15, 2011Z-2558-2011The values reported by the chloride sensors may report erroneous values below specification on aqueous controls and potentially with patient samples of whole blood and serum/plasma.Terminated
Apr 20, 2009Z-1083-2010Outer box labeling shows an incorrect storage temperature range.Terminated
Jul 18, 2007Z-0119-2008Incorrect results/stability failure; Actual device stability studies show sodium and chloride sensors age differently than anticipated when stored at the high end of labeled storage, therefore subsequent measurements…Terminated

Record

Type
Traditional
Received
Jul 27, 2005
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1170

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.