Capiox RX15 Hollow, Fiber Oxygenator With/Without Hardshell Reservoir
Oxygenator, Cardiopulmonary Bypass. 510(k) K051997, from Terumo Cardiovascular Systems Corp. (Elkton, MD). Product code DTZ, class 2.
As of Nov 21, 2026, K051997 is an FDA 510(k) held by Terumo Cardiovascular Systems Corp., substantially equivalent on Sep 15, 2005 after 52 days, with 1 recall tied to it in FDA's recall database.
- Review time: 52 days
- Recalls: 1
- Cited as a predicate by: 5 devices
The Capiox RX15 is intended to be used during open-heart surgical procedures requiring cardiopulmonary bypass for period up to 6 hours. The Capiox RX15 is for use with patients when required blood flow rate will not exceed 5 L/min (4 L/min if using the CX*RX15RW30 product code or CX*RX15RE30 Product code).
- Capiox SX10 Hollow Fiber Oxygenator With Detachable Hardshell Venous ReservoirK960074 · Terumo Medical Corp. · Oct 16, 1996
- Modification To Capiox SX10 Hardshell ReservoirK030449 · Terumo Corp. · Mar 5, 2003
- Minimax Plus Hollow Fiber Oxygenator (With Or Without Medtronic/Carmeda Bioactive SurfaceK933586 · Medtronic Vascular · Feb 7, 1994
- Quadrox-I Small Adult With and Without Intergrated Arterial Filter With Softline Coating, Models Hmo 50000, Hmo 51000K090689 · Maquet Cardiopulmonary, AG · Oct 29, 2009
- Medos Hilite Pediatric Oxygenator, Models 2800, 2400 LTK090450 · Gish Biomedical, Inc. · Sep 18, 2009
- Terumo Rocsafe Hybrid Perfusion SystemK082321 · Terumo Cardiovascular Systems Corp. · Feb 13, 2009
- Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterK071494 · Terumo Corp. · Jul 16, 2007
- Capiox RX Oxygenator With Or Without ReservoirK062381 · Terumo Cardiovascular Systems Corp. · Aug 24, 2006
| Started | Recall | Status |
|---|---|---|
| Apr 12, 2012 | Z-1523-2012Loose connections (i.e., connections between PVC circuit tubing and Oxygenator ports) may result in leaks and patient blood loss | Terminated |
- HORIZON 8AF (Integrated OXY MODULE 8AF and Horizon CVR) (US5305)K261438 · Eurosets S.R.L · Sep 14, 2026
- Lifemotion Disposable Membrane OxygenatorK253838 · Lifemotion Medical Technology Co., Ltd. · Mar 16, 2026
- Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum Perfusion Blood Oxygenator ECC VT200-E3 (VT200-E3);…K240934 · Spectrum Medical S.R.L. · Mar 12, 2026
- Horizon AF Plus System (Integrated A.L. One AF Plus OXYGENATOR and Horizon CVR) (US5300); A.L. One AF Plus OXYGENATOR…K250821 · Eurosets S.R.L · Dec 12, 2025
- Trilly Pediatric Oxygenator with Integrated Arterial Filter (US5232)K242503 · Eurosets S.R.L · Nov 24, 2025
- Quantum Perfusion Blood Oxygenator VT75-E1, Quantum Perfusion Blood Oxygenator VT75-E2K231982 · Qura S.R.L · Sep 12, 2024
- Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811); Affinity Fusion Oxygenator…K240666 · Medtronic Perfusion Systems · Jul 10, 2024
- Affinity NT Oxygenator and Uncoated Cardiotomy/ Venous Reservoir (541); Affinity NT Oxygenator with Balance Biosurface…K241352 · Medtronic, Inc. · Jun 12, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 25, 2005
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4350
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files