Arca
Skip to content
Tools
Tools

Revolution Xr/D Digital Radiographic Imaging System With Tomosynthesis

System, X-Ray, Stationary. 510(k) K051967, from GE Healthcare (Waukesha, WI). Product code KPR, class 2.

As of Nov 20, 2026, K051967 is an FDA 510(k) held by GE Healthcare, substantially equivalent on Aug 9, 2005 after 20 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 20 days
  • Recalls: 9
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Revolution XR/d Digital Radiographic Imaging System is indicated for use in generating Tomographic images of human anatomy. This device is not intended for mammographic applications.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K051967, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 26, 2011Z-1065-2012GE Healthcare has become aware of a potential issue associated with the hand grip that is clamped to the table top of your Revolution XRD X-ray systems that may impact patient safety. The hand grip accessory that is…Terminated
Aug 13, 2009Z-0528-2010Certain GE Radiographic fixed X-ray tables (Proteus XR/a, Revolution XR/d, and Definium 8000) were found to have tabletops that move unexpectedly in a bi-directional mode. In rare cases, a malfunction may occur, which…Terminated
Mar 7, 2008Z-1822-2008Loss of image data: Software anomaly in the Processing Software on the Definium 8000 system may impact patient safety when using the VolumeRAD advanced application (option). The slice visualization of VolumeRAD exams…Terminated
Mar 7, 2008Z-1882-2008Radiation dose reading greater than actual dose: Error in the generator software on the GE Healthcare Revolution XR/d configured with HP 8200 PC consoles and all Definium 8000 systems may impact patient safety.…Terminated
Feb 14, 2008Z-0002-2009GE Healthcare has recently become aware of a potential risk of observing a blank image after exposure on Revolution XR/d systems equipped with HP 8200 PC user consoles that may impact patient safety. It was reported…Terminated
Feb 14, 2008Z-0003-2009GE Healthcare has recently become aware of an error in the generator software on the Revolution XR/d configured with HP 8200 PC consoles that reports a mAs reading that is higher than is actually exposed to patients.Terminated
Jan 2, 2008Z-1883-2008Not properly labeling the manual switch for disabling positive beam limitation (PBL) located on the overhead tube support user interface as requiredTerminated
Aug 13, 2007Z-1476-2008Necessary certification labels are missing from the x-ray contol.Terminated
May 31, 2006Z-1788-2012An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers…Terminated

Other KPR clearances

The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device

Record

Type
Traditional
Received
Jul 20, 2005
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.