Arca
Skip to content
Tools
Tools

Cobalt Bone Cement

Bone Cement. 510(k) K051496, from Biomet, Inc. (Warsaw, IN). Product code LOD, class 2.

As of Nov 19, 2026, K051496 is an FDA 510(k) held by Biomet, Inc., substantially equivalent on Aug 4, 2005 after 59 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 59 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Cobalt™ Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K051496, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 26, 2017Z-0067-2018, Z-0064-2018Loss of the seal on the sterile Tyvek packaging used with this Cobalt Bone Cement.Open
Apr 13, 2017Z-2056-2017Six complaints were filed regarding loss of the Zimmer Biomet applied seal of a sterile Tyvek packaging. A breach in the packaging could lead to loss of sterility of the device.Terminated
Jul 26, 2016Z-2606-2016The outer packaging was mislabeled on the box indicating "Cobalt HV with Gentamicin". The bone cement does not contain antibiotics.Terminated

Other LOD clearances

The newest 510(k)s for the same product code; their median review was 204 days; see every LOD device

Record

Type
Traditional
Received
Jun 6, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3027

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files