Humeral Head Plate With Angular Stability

Plate, Fixation, Bone. 510(k) K051412, from I.T.S. Implantat-Technologie-Systeme GmbH (Prior Lake, MN). Product code HRS, class 2.

As of Nov 18, 2026, K051412 is an FDA 510(k) held by I.T.S. Implantat-Technologie-Systeme GmbH, substantially equivalent on Jul 11, 2005 after 41 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Jul 11, 2005
  • Review time: 41 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The I.T.S. Humeral Head Plate with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal humerus in the shoulder. Indications for Use include fracture and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K051412, newest first; each opens FDA's recall record
StartedReasonStatus
Jan 23, 2026Z-1594-2026Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).Open, Classified

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
May 31, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files