O-Arm Imaging System
Image-Intensified Fluoroscopic X-Ray System, Mobile. 510(k) K050996, from Breakaway Imaging, LLC (Littleton, MA). Product code OXO, class 2.
As of Nov 19, 2026, K050996 is an FDA 510(k) held by Breakaway Imaging, LLC, substantially equivalent on May 5, 2005 after 15 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 4
- Cited as a predicate by: 3 devices
The O-Arm TM Imaging System is designed for 2D Fluoroscopic and 3D imaging for intraoperative applications in surgical theaters, particularly for orthopedic applications. The O-Arm TM Imaging System is compatible with certain Image Guided Surgical Systems.
- Siremobil Iso-C 3DK032280 · Siemens AG · Sep 10, 2003
- Siremobil Iso-C 3DK003266 · Siemens Medical Solutions USA, Inc. · Dec 1, 2000
- O-Arm O2 Imaging SystemK151000 · Medtronic, Inc. · Aug 6, 2015
- C-InsightK081672 · Mazor Surgical Technologies , Ltd. · Aug 15, 2008
- O-Arm Imaging SystemK060344 · Breakaway Imaging, LLC · Mar 29, 2006
| May 10, 2010 | Z-1998-2010The possibility that the real time image display can freeze up during use of a 2D continuous fluoroscopy. | Terminated |
| Jun 30, 2008 | Z-2034-2008Failure to of affected units to comply with EER/AKR limits due to misinterpretation of the measurement requirements specified in 21 CFR 1020.32 (d) (3) (iii) | Terminated |
| Nov 27, 2007 | Z-0624-2009Navigation accuracy problem: Accuracy may fall outside of specification when used in combination with the Medtronic Synergy Experience StealthStation System Spine Software with the O-Arm Intraoperative Imaging System… | Terminated |
| Oct 17, 2007 | Z-0294-2008Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is… | Terminated |
- Digital C-ARM with 3DK253296 · Allengers Medical Systems Limited · Jun 5, 2026
- Orthoscan TAU MVP Mini C-Arm SystemK252579 · Ziehm-Orthoscan, Inc. · Jan 14, 2026
- OEC One CFDK253269 · Ge Hualun Medical Systems Co. , Ltd. · Nov 26, 2025
- Dual-Mode Mobile C-Arm (Geelin500A, Geelin500M)K251004 · Hefei Chimed Intelligent Machine Co., Ltd. · Nov 6, 2025
- Orthoscan TAU Mini C-ArmK250587 · Ziehm-Orthoscan, Inc. · Jul 2, 2025
- Orthoscan VERSA Mini C-ArmK243452 · Ziehm-Orthoscan, Inc. · Jan 14, 2025
- OEC One ASDK240828 · Ge Hualun Medical Systems Co. , Ltd. · Dec 27, 2024
- Oec 3DK233669 · Ge Oec Medical Systems, Inc. · Mar 28, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated, third-party review
- Received
- Apr 20, 2005
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files