Neurological Stereotaxic Instrument. 510(k) K050438, from Medtronic Surgical Navigation Technologies (Louisville, CO). Product code HAW, class 2.

As of Nov 17, 2026, K050438 is an FDA 510(k) held by Medtronic Surgical Navigation Technologies, substantially equivalent on Jun 2, 2005 after 100 days, with 11 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 2, 2005
  • Review time: 100 days
  • Recalls: 11
  • Cited as a predicate by: 21 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

This submission does not change the indications for use for the StealthStation System. The StealthStation® System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.

Cited as a Predicate

21 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K050438, newest first; each opens FDA's recall record
May 10, 2017Z-2746-2017Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated…Terminated
Jul 18, 2016Z-2313-2016Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.Terminated
Jul 18, 2016Z-2318-2016Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.Terminated
Sep 14, 2015Z-2824-2015Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.Terminated
Nov 13, 2012Z-0441-2013Medtronic is recalling certain Vertex Max Drill Tubes which are used with the StealthStation System due to a defect related to the internal threading for component set up.Terminated
Jun 2, 2011Z-2781-2011Software allows incorrect patient or patient history to be displayed.Terminated
Dec 22, 2010Z-1558-2011Position Sensor Unit may generate tracking errors or may stop tracking immediately.Terminated
Dec 15, 2010Z-1042-2011Incorrectly manufactured drill tube could cause incorrect screw placement during spinal surgery.Terminated
Nov 5, 2010Z-0715-2011Electronic surgical navigation aid may lose tracking ability or stop working.Terminated
Sep 24, 2010Z-0145-2011Drill guide handle may detach unintentionally during use.Terminated
Jun 14, 2010Z-2029-2010Excessive wear at the power cord retractor or incorrect assembly of the power plug may cause sparking, intermittent loss of power, or electrical shock.Terminated

Other HAW Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Special
Received
Feb 22, 2005
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.