Diver C.E. Catheter
Aspiration Thrombectomy Catheter. 510(k) K050276, from Invatec Innovative Technologies, S.R.L. (Plymouth, MN). Product code QEZ, class 2.
As of Nov 17, 2026, K050276 is an FDA 510(k) held by Invatec Innovative Technologies, S.R.L., substantially equivalent on Mar 7, 2005 after 28 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Mar 7, 2005
- Review time: 28 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe DIVER C.E. Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Vascular Solutions Pronto Extraction CatheterK032763 · Vascular Solutions, Inc. · Dec 19, 2003
- Submarine Plus Percutaneous Transluminal Angioplasty (Pta) Catheter, Model SuvxxxxxxxxxK042537 · Invatec Innovative Technologies · Nov 8, 2004
- Amphirion Deep 0.014 Otw Pta Balloon CatheterK042624 · Invatec Innovative Technologies, S.R.L. · Jan 10, 2005
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Qxt Extraction CatheterK071727 · Vascular Solutions, Inc. · Sep 28, 2007
- Quickcat Extraction CatheterK060092 · Kensey Nash Corp. · Mar 9, 2006
- Diver C.E. CatheterK051917 · Invatec Innovative Technologies, S.R.L. · Aug 8, 2005
Recalls
Recalls FDA ties to K050276, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Mar 4, 2009 | Z-1517-2013Devices labeled for an intended use was not included in the existing 510(k) or PMA, in that, the Indications for use was modified. | Terminated |
Other QEZ Clearances
The newest 510(k)s for the same product code; their median review was 108 days; see every QEZ device- FLOWRUNNER Aspiration SystemK262864 · Expanse Medical, Inc. · Oct 26, 2026
- Mayumi Thrombectomy SystemK254035 · MicroVention, Inc., d/b/a Terumo Neuro · Sep 3, 2026
- Powered Aventus Thrombectomy SystemK261105 · Inquis Medical · Jul 27, 2026
- Aventus Thrombectomy SystemK262030 · Inquis Medical · Jul 15, 2026
- Liberant RX Aspiration CatheterK253589 · Medtronic Interventional Vascular, Inc. · Apr 23, 2026
- FLOWRUNNER Aspiration SystemK254208 · Expanse Medical, Inc. · Mar 20, 2026
- Aventus Thrombectomy SystemK253925 · Inquis Medical · Jan 15, 2026
- AlphaVac Multipurpose Mechanical Aspiration System F1885K252509 · AngioDynamics, Inc. · Nov 24, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 7, 2005
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5150
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files