Arca
Skip to content
Tools
Tools

Ebi Trochanteric Nail System

Rod, Fixation, Intramedullary And Accessories. 510(k) K050118, from Ebi, L.P. (Parsippany, NJ). Product code HSB, class 2.

As of Nov 18, 2026, K050118 is an FDA 510(k) held by Ebi, L.P., substantially equivalent on Feb 16, 2005 after 29 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 16, 2005
  • Review time: 29 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The EBI$^{®}$ Trochanteric Nail is indicated for the treatment of fractures of the femur including: non-comminuted and comminuted midshaft fracture, subtrochanteric fracture, distal third fracture, combination fractures of the shaft and neck, intertrochanteric fracture, combination intertrochanteric and subtrochanteric fractures.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K050118, newest first; each opens FDA's recall record
Apr 1, 2016Z-1819-2016Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is…Terminated
Mar 30, 2012Z-1478-2012The Trochanteric Nail, lot 486130, was mislabeled as a right nail orientation but the contents are actually a left nail orientationTerminated
Jun 24, 2005Z-0012-06Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.Terminated

Other HSB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device

Record

Type
Traditional
Received
Jan 18, 2005
Review panel
Orthopedic
Regulation
21 CFR 888.3020

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.