Model 2500A Palmsat Pulse Oximeter
Oximeter. 510(k) K050056, from Nonin Medical, Inc. (Plymouth, MN). Product code DQA, class 2.
As of Nov 17, 2026, K050056 is an FDA 510(k) held by Nonin Medical, Inc., substantially equivalent on Jun 21, 2005 after 161 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Jun 21, 2005
- Review time: 161 days
- Recalls: 1
- Cited as a predicate by: 6 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Nonin® Model 2500A Pulse Oximeter with Alarms is indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO₂) and pulse rate for adult, pediatric, and neonatal patients. It is intended for spot-checking and / or continuous monitoring of patients during motion and no motion conditions, and for patients who are well or poorly perfused.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Model 8500, Hand Held Pulse Oximeter, Model 8500K001930 · Nonin Medical, Inc. · Jul 20, 2000
- Model 2500, Palmsat Hand Held Pulse Oximeter, Model 2500K002690 · Nonin Medical, Inc. · Oct 11, 2000
Cited as a Predicate
6 later 510(k) summaries name this device as a predicate- Mini-MedicK113165 · Athena Gtx · Feb 2, 2012
- Single-Patient Use Disposable Sensor SeriesK093853 · Nonin Medical, Inc. · Jun 4, 2010
- Single-Patient Use Disposable SensorK093306 · Nonin Medical, Inc. · Mar 4, 2010
- Reusable Pulse Oximeter Sensor, Model 8000 SeriesK092101 · Nonin Medical, Inc. · Oct 21, 2009
- Mediaid Pulse Oximeter M30 And M34K071610 · Mediaid, Inc. · Nov 5, 2007
- Digital Pulse Oximeter, Model 7500K071285 · Nonin Medical, Inc. · Jul 12, 2007
Recalls
Recalls FDA ties to K050056, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Oct 11, 2007 | Z-0843-2008Incorrect Label: The label on the back of the device may show Model 2500A rather than Model 2500. | Terminated |
Other DQA Clearances
The newest 510(k)s for the same product code; their median review was 207 days; see every DQA device- Ninenovo Blood Oxygen Ring…K260490 · Shenzhen Ninenovo Technology Limited · Oct 2, 2026
- Reusable Adult SpO2 Clip Sensor (3m) (M1196A); Reusable Adult SpO2 Clip Sensor (0.9m)…K252821 · Philips Medizin Systeme Böblingen GmbH · May 29, 2026
- Pulse Oximeter (PO-A2AO, PO-A2AT, PO-A3AO, PO-H1AO, PO-B1AO, PO-C5AO, PO-C5AT, PO-C6AO,…K252655 · Shenzhen Imdk Medical Technology Co., Ltd. · May 11, 2026
- Nasal Alar SpO2 Sensor (989803205381); Nasal Alar SpO2 Sensor (989803205391); Nasal Alar…K253887 · Philips Medizin Systeme · May 1, 2026
- Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)K260931 · Unimed Medical Supplies, Inc. · Apr 14, 2026
- AViTA Pulse Oximeter (SP61)K252448 · Avita Corporation · Feb 27, 2026
- YUWELL® Finger Pulse Oximeter (YX105, YX106, YX110, YX310)K252805 · Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · Feb 17, 2026
- Reusable SPO2 Sensor (BSA307-47; BSA109-50; BSA109-31O)K253109 · Shenzhen Best Electronics Co., Ltd. · Jan 7, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 11, 2005
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2700
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files