Liberty System
Electrocardiograph. 510(k) K043074, from Inovise Medical, Inc. (Portland, OR). Product code DPS, class 2.
As of Nov 18, 2026, K043074 is an FDA 510(k) held by Inovise Medical, Inc., substantially equivalent on Feb 22, 2005 after 106 days, with 1 recall tied to it in FDA's recall database.
- Review time: 106 days
- Recalls: 1
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Liberty System, when used with Audicor Sensors on the chest wall, is intended for use in acquiring, analyzing and reporting ECG and heart sound data and to provide interpretation of the data in an integrated COR report for consideration by physicians.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Audicor Upgrade SystemK032145 · Inovise Medical, Inc. · Oct 31, 2003
- Eli 200+ AudicorK031182 · Mortara Instrument, Inc. · Jul 25, 2003
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Audicor 200 System, Models 200D and 200SK070136 · Inovise Medical, Inc. · Apr 13, 2007
- Extended Measurements SystemK051450 · Inovise Medical, Inc. · Nov 10, 2005
Recalls
Recalls FDA ties to K043074, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Jan 12, 2003 | Z-0565-06Use of the AUDICOR 200 D in-line with a defibrillator/monitor resulted in interruption of pacing therapy to a patient when the AUDICOR 200 D stopped working due to low battery power. | Terminated |
Other DPS clearances
The newest 510(k)s for the same product code; their median review was 241 days; see every DPS device- Kinetic Software LibraryK260640 · Monebo Technologies, Inc. · Oct 23, 2026
- MyoVista wavECG (DG-200-B15-003)K260005 · HeartSciences, Inc. · Oct 5, 2026
- QCA 10 Electrocardiograph (QCARD-10); QCA 10W Electrocardiograph (QCARD-10W)K260250 · QT Medical, Inc. · Sep 29, 2026
- SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)K252463 · Edan Instruments, Inc. · Jul 2, 2026
- MAC 7 Resting ECG Analysis SystemK251670 · Ge Medical Systems Information Technologies, Inc. · Jan 26, 2026
- AccurECG Analysis System (v2.0)K252361 · Accurkardia, Inc. · Dec 22, 2025
- Withings BeamO (SCT02)K252474 · Withings · Nov 10, 2025
- Cardiologs Holter PlatformK250569 · Philips France Commercial · Aug 6, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 8, 2004
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2340
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files