Arca
Skip to content
Tools
Tools

Lysus Infusion System

Mechanical Thrombolysis Catheter. 510(k) K042456, from Ekos Corp. (Bothell, WA). Product code QEY, class 2.

As of Nov 19, 2026, K042456 is an FDA 510(k) held by Ekos Corp., substantially equivalent on Sep 23, 2004 after 13 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 13 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Lysus® Infusion System is intended for the controlled and selective infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K042456, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 18, 2005Z-0175-06Potential for a faulty thermal channel offset to be reported to the software causing faulty lower temperature values to be reported which allow a power increase.Terminated
Nov 22, 2004Z-0111-06During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In…Terminated
Sep 21, 2004Z-0110-06The 100V board in the catheter interface cable pod failed during use. The PT-3 control unit would not recognize the catheter interface cable.Terminated

Other QEY clearances

The newest 510(k)s for the same product code; their median review was 151 days; see every QEY device

Record

Type
Special
Received
Sep 10, 2004
Review panel
Cardiovascular
Regulation
21 CFR 870.5150

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.