Fabius Gs/Fabius Tiro Anesthesia System

Gas-Machine, Anesthesia. 510(k) K042419, from Draeger Medical, Inc. (Telford, PA). Product code BSZ, class 2.

As of Nov 20, 2026, K042419 is an FDA 510(k) held by Draeger Medical, Inc., substantially equivalent on Oct 8, 2004 after 31 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 31 days
  • Recalls: 11
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Fabius GS/Fabius Tiro is indicated as a continuous flow anesthesia system. The Fabius GS/Fabius Tiro can be used for spontaneous, manually assisted, automatic, pressure support, or synchronized mandatory intermittent ventilation, delivery of gases and anesthetic vapor, and monitoring oxygen concentration, breathing pressure and respiratory volume of patients during anesthesia.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Fabius MriK072884 · Draeger Medical AG & Co. KG · Jul 30, 2008

Recalls

Recalls FDA ties to K042419, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 24, 2018Z-2250-2018, Z-2251-2018The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient.…Terminated
Mar 1, 2018Z-1482-2018, Z-1484-2018, Z-1483-2018The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.Terminated
Aug 14, 2013Z-0008-2014During final testing of the Draeger Fabius GS Premium, Fabius OS, and Fabius Tiro anesthesia machines, some units did not pass the high voltage test and the minimum clearance between an electrical component and the unit…Terminated
Aug 26, 2010Z-2479-2010, Z-2478-2010potential failures to Electri-Cord Manufacturing AC power cordTerminated
Jul 7, 2008Z-2338-2008The parent company became aware of instances where the lower side rail of the frame of the machine broke inwards when being moved across a threshold; resulting in the machine being unstable.Terminated
Feb 5, 2008Z-1315-2008Detachment of Component: The chassis on some machines were manufactured out of specifications. As a result, the caster may break loose from the chassis.Terminated
Oct 3, 2006Z-0048-2007One of four casters may break loose from chassisTerminated

Other BSZ clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Traditional
Received
Sep 7, 2004
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files