Somatom Spirit
System, X-Ray, Tomography, Computed. 510(k) K042328, from Siemens Medical Systems, Inc. (Mavern, PA). Product code JAK, class 2.
As of Nov 20, 2026, K042328 is an FDA 510(k) held by Siemens Medical Systems, Inc., substantially equivalent on Sep 14, 2004 after 18 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 18 days
- Recalls: 2
- Cited as a predicate by: 1 device
- Somatom Emotion MSK003014 · Siemens Medical Solutions USA, Inc. · Oct 12, 2000
- Hispeed Nx/I CT Scanner SystemK991716 · Ge Medical Systems, Inc. · Jun 18, 1999
- Somatom Scope/Somatom Scope PowerK140912 · Siemens Medical Solutions USA, Inc. · Sep 15, 2014
| Feb 15, 2018 | Z-1451-2018to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical… | Terminated |
| Mar 21, 2014 | Z-1721-2014The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury. | Terminated |
- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 27, 2004
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files