Arca
Skip to content
Tools
Tools

Robocouch Patient Support System

Couch, Radiation Therapy, Powered. 510(k) K042146, from Accuray Incorporated (Sunnyvale, CA). Product code JAI, class 2.

As of Nov 20, 2026, K042146 is an FDA 510(k) held by Accuray Incorporated, substantially equivalent on Sep 17, 2004 after 39 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 39 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The RoboCouch Patient Support System is indicated for the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K042146, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 27, 2013Z-1155-2013A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.Terminated
Jul 21, 2009Z-0853-2011Mechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.Terminated
Oct 30, 2008Z-0856-2009Product may not be tensioned properly, potentially causing unexpected rotation or descent.Terminated

Other JAI clearances

The newest 510(k)s for the same product code; see every JAI device

Record

Type
Traditional
Received
Aug 9, 2004
Review panel
Radiology
Regulation
21 CFR 892.5770

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.