Synchron Systems Phosphorus (Phs) Reagent
Phosphomolybdate (Colorimetric), Inorganic Phosphorus. 510(k) K041643, from Beckman Coulter, Inc. (Brea, CA). Product code CEO, class 1.
As of Nov 19, 2026, K041643 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Aug 18, 2004 after 62 days, with no recalls tied to it in FDA's recall database.
- Review time: 62 days
- Recalls: 0
PHS reagent, in conjunction with SYNCHRON® Systems and SYNCHRON Multi Calibrator, is intended for use in the quantitative determination of inorganic phosphorus in human serum, plasma, or urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.
- Synchron CX4/CX5 Clinical Chemistry Analyzer ReageK883181 · Beckman Instruments, Inc. · Sep 19, 1988
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 17, 2004
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1580
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files