Arca
Skip to content
Tools
Tools

Precision Rxi Analog X-Ray System

Interventional Fluoroscopic X-Ray System. 510(k) K041605, from General Medical Merate S.P.A (Seriate (Bg), IT). Product code OWB, class 2.

As of Nov 22, 2026, K041605 is an FDA 510(k) held by General Medical Merate S.P.A, substantially equivalent on Jun 30, 2004 after 15 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 15 days
  • Recalls: 5

Recalls

Recalls FDA ties to K041605, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 1, 2008Z-2145-2008The Precision RXi collimator did not contain a label identifying the device as being certified to comply with applicable requirements of the x-ray performance standard.Terminated
Dec 18, 2007Z-1565-2008, Z-1563-2008, Z-1564-2008GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the…Terminated
Nov 1, 2006Z-1485-06Automatic exposure control (AEC) automatically resets allowing another exposure without requiring manual reset as specified in 21 CFR 1020.31 (a)(3)(iv)Terminated

Other OWB clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional, third-party review
Received
Jun 15, 2004
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.