Zimmer Trabecular Metal Humeral Stem
Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented. 510(k) K041549, from Zimmer, Inc. (Warsaw, IN). Product code KWT, class 2.
As of Nov 18, 2026, K041549 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Sep 1, 2004 after 84 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 1, 2004
- Review time: 84 days
- Recalls: 8
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFProsthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; ununited humeral head fractures of long duration; irreducible 3-and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- New Zimmer Shoulder SystemK982981 · Zimmer, Inc. · Dec 17, 1998
- Bio-Modular Shoulder SystemK992119 · Biomet, Inc. · Sep 13, 1999
Recalls
Recalls FDA ties to K041549, newest first; each opens FDA's recall record| Jan 10, 2014 | Z-1092-2014This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of… | Terminated |
| Sep 10, 2010 | Z-0576-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating… | Terminated |
| Sep 10, 2010 | Z-0573-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating… | Terminated |
| Sep 10, 2010 | Z-0575-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating… | Terminated |
| Sep 10, 2010 | Z-0570-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating… | Terminated |
| Sep 10, 2010 | Z-0574-2011The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating… | Terminated |
| Sep 17, 2009 | Z-0178-2010Some of the devices may possess a taper which will not mate with the humeral head, as intended. | Terminated |
| Sep 17, 2009 | Z-0177-2010Some of the devices may possess a taper which will not mate with the humeral head, as intended. | Terminated |
Other KWT Clearances
The newest 510(k)s for the same product code; their median review was 232 days; see every KWT device- N22 Offset Humeral HeadK262173 · Shoulder Innovations, Inc. · Sep 8, 2026
- Equinoxe® Humeral Reconstruction Prosthesis; Equinoxe® Stemless Shoulder ImplantsK250713 · Exactech, Inc. · Dec 1, 2025
- Tornier Humeral Reconstruction System Max (Tornier HRS Max)K251686 · Tornier, Inc. · Oct 30, 2025
- AGILON® XO Shoulder Replacement SystemK241944 · Implantcast GmbH · Mar 10, 2025
- InSet Total Shoulder SystemK241817 · Shoulder Innovations, Inc. · Feb 11, 2025
- AGILON® XO Shoulder Replacement SystemK231657 · Implantcast GmbH · Feb 16, 2024
- Comprehensive Segmental Revision System(SRS)K223631 · Biomet Orthopedics · Jul 12, 2023
- AGILON XO Shoulder SystemK222482 · Implantcast GmbH · Sep 28, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 9, 2004
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files