Sentec Digital Monitor Model #Sdm

Monitor, Carbon-Dioxide, Cutaneous. 510(k) K041548, from Sentec, Inc. (Egale, WI). Product code LKD, class 2.

As of Nov 18, 2026, K041548 is an FDA 510(k) held by Sentec, Inc., substantially equivalent on Aug 9, 2004 after 61 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 61 days
  • Recalls: 9
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SenTec Digital Monitor System (comprising the SenTec Digital Monitor, V-Signā„¢ Sensor and Accessories) is indicated for continuous non-invasive patient monitoring. The SenTec Digital Monitor System is indicated for use in hospitals, hospital-type facilities, intra-hospital transport environments, and, if under clinical supervision, home environments. The SenTec Digital Monitor System is for

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K041548, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 25, 2023Z-2592-2023During rework of components during the manufacturing process, an additional piece of material may have been left in the device, which may result in an impact on measurement performance.Open
May 2, 2023Z-1980-2023The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.Open
Dec 10, 2020Z-1134-2021Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards.Terminated
May 2, 2019Z-2153-2019, Z-2154-2019, Z-2155-2019, Z-2156-2019, Z-2152-2019There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.Terminated
Sep 15, 2016Z-0088-2017Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the…Terminated

Other LKD clearances

The newest 510(k)s for the same product code; see every LKD device

Record

Type
Traditional
Received
Jun 9, 2004
Review panel
Anesthesiology
Regulation
21 CFR 868.2480

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.