Arca
Skip to content
Tools
Tools

Liquichek Torch Plus Igm Control

Multi-Analyte Controls, All Kinds (Assayed). 510(k) K040822, from Bio-Rad Laboratories, Inc. (Irvine, CA). Product code JJY, class 1.

As of Nov 27, 2026, K040822 is an FDA 510(k) held by Bio-Rad Laboratories, Inc., substantially equivalent on May 19, 2004 after 50 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 50 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Liquichek ToRCH Plus IgM Control is intended for use as an unassayed quality control serum to monitor precision of IgM laboratory testing procedures for the analytes listed in this package insert. This product is not intended for use in blood donor screening assays.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Virotrol TorchK942295 · Blackhawk Biosystems, Inc. · Oct 7, 1994

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K040822, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Apr 15, 2013Z-1330-2013The Liquichek" ToRCH Plus IgM Control - Positive, contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this…Terminated

Record

Type
Traditional
Received
Mar 30, 2004
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1660

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.