Arca
Skip to content
Tools
Tools

Bardex I.C. Latex Foley Catheter

Catheter, Urological (Antimicrobial) And Accessories. 510(k) K040658, from C.R. Bard, Inc. (Covington, GA). Product code MJC, class 2.

As of Nov 21, 2026, K040658 is an FDA 510(k) held by C.R. Bard, Inc., substantially equivalent on Sep 29, 2004 after 201 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 201 days
  • Recalls: 9
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Bardex® I.C. latex Foley catheter is indicated for use in the drainage and/or collection and/or measurement of urine. Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as nephrostomy tract.

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K040658, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 25, 2026Z-3116-2026Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.Open
Jul 17, 2025Z-2591-2025, Z-2589-2025, Z-2590-2025Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating,…Open
May 9, 2024Z-2154-2024Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature monitoring devices.Open
Aug 5, 2021Z-2525-2021, Z-2523-2021, Z-2527-2021There is potential for packaging defects that may impact the sterile barrier.Open
Apr 15, 2020Z-1970-2020There is potential for a white paper-type residue attached to the catheter.Terminated

Other MJC clearances

The newest 510(k)s for the same product code; see every MJC device

Record

Type
Traditional
Received
Mar 12, 2004
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5130

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files