Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component). 510(k) K040627, from DePuy Orthopaedics, Inc. (Warsaw, IN). Product code KWA, class 3.
As of Nov 18, 2026, K040627 is an FDA 510(k) held by DePuy Orthopaedics, Inc., substantially equivalent on Aug 5, 2005 after 514 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 514 days
- Recalls: 25+
- Cited as a predicate by: 6 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe DePuy ASR™ Modular Acetabular Cup System is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Pinnacle Metal-On-Metal Acetabular Cup LinersK002883 · DePuy Orthopaedics, Inc. · Oct 13, 2000
- Depuy Pinnacle Metal-On-Metal Acetabular Cup LinersK003523 · DePuy Orthopaedics, Inc. · Dec 13, 2000
- Metal Transcend Articulation System (Larger Sizes)K021349 · Wrightmedicaltechnologyinc · Jul 1, 2002
- Ultima*Unipolar Head and Adapter SleevesK965156 · Johnson & Johnson Professionals, Inc. · Jan 24, 1997
Cited as a predicate
6 later 510(k) summaries name this device as a predicate- Depuy Asr XL Modular Acetabular Cup SystemK080991 · DePuy Orthopaedics, Inc. · Jul 2, 2008
- Ear Thermometer, Models ST66 and ST67K080981 · Mesure Technology Co., Ltd. · May 2, 2008
- Depuy Asr 300 Acetabular Cup SystemK073413 · DePuy Orthopaedics, Inc. · Jan 30, 2008
- Depuy Asr Taper Sleeve AdapterK070359 · DePuy Orthopaedics, Inc. · Mar 6, 2007
- Depuy Pinnacle Metal-On-Metal Acetabular Cup LinersK062426 · DePuy Orthopaedics, Inc. · Dec 15, 2006
- M2A Magnum 12/14 Taper Inserts and One-Piece Modular HeadsK061423 · Biomet Manufacturing Corp · Jul 27, 2006
Recalls
Recalls FDA ties to K040627, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Started | Recall | Status |
|---|---|---|
| Aug 23, 2010 | Z-1791-2011, Z-1762-2011, Z-1751-2011, Z-1753-2011, Z-1779-2011, Z-1752-2011, Z-1787-2011, Z-1777-2011, Z-1790-2011, Z-1760-2011, Z-1759-2011, Z-1786-2011, Z-1757-2011, Z-1758-2011, Z-1755-2011, Z-1794-2011, Z-1780-2011, Z-1785-2011, Z-1788-2011, Z-1782-2011, Z-1756-2011, Z-1784-2011, Z-1776-2011, Z-1789-2011, Z-1750-2011As part of our ongoing post-market surveillance of all products, DePuy is continually evaluating data from a variety of sources including national joint replacement registries, published literature, company sponsored… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 9, 2004
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3330
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files