Arca
Skip to content

Reflex Hybrid Anterior Cervical Plate System

Appliance, Fixation, Spinal Intervertebral Body. 510(k) K040261, from Stryker Spine (Allendale, NJ). Product code KWQ, class 2.

As of Nov 17, 2026, K040261 is an FDA 510(k) held by Stryker Spine, substantially equivalent on Apr 16, 2004 after 72 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 16, 2004
  • Review time: 72 days
  • Recalls: 4
  • Cited as a predicate by: 8 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Reflex™ Hybrid ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: - Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc…

Cited as a Predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K040261, newest first; each opens FDA's recall record
Aug 18, 2008Z-0217-2009The Reflex Hybrid 4.0 x 14 mm Variable, Self Tapping Bone Screw, Lot NSZ, was determined to have been anodized with the incorrect color for that screw.Terminated
Jun 14, 2006Z-1958-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.Terminated
Jun 14, 2006Z-1960-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.Terminated
Jun 14, 2006Z-1959-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used.Terminated

Other KWQ Clearances

The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device

Record

Type
Special
Received
Feb 4, 2004
Review panel
Orthopedic
Regulation
21 CFR 888.3060

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.