Reflex Hybrid Anterior Cervical Plate System
Appliance, Fixation, Spinal Intervertebral Body. 510(k) K040261, from Stryker Spine (Allendale, NJ). Product code KWQ, class 2.
As of Nov 17, 2026, K040261 is an FDA 510(k) held by Stryker Spine, substantially equivalent on Apr 16, 2004 after 72 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 16, 2004
- Review time: 72 days
- Recalls: 4
- Cited as a predicate by: 8 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Reflex™ Hybrid ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: - Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc…
Cited as a Predicate
8 later 510(k) summaries name this device as a predicate- CarboClear® Cervical Plate SystemK200883 · CarboFix Orthopedics , Ltd. · Jun 4, 2020
- Ocata Anterior Cervical SystemK170342 · Kalitec Direct, LLC · May 9, 2017
- Aviator Anterior Cervical Plate (ACP) SystemK142237 · Stryker Corporation · Jan 26, 2015
- Stryker Spine Nulock Anterior Cervical Plating SystemK083562 · Stryker Corp. · Mar 25, 2009
- Stryker Spine Reflex Translational Anterior Cervical Plating SystemK083020 · Stryker Corp. · Mar 26, 2009
- Zero Profile Anterior Cervical Plate (Acp) SystemK063430 · Stryker Spine · Mar 22, 2007
- Reflex Hybrid Acp SystemK062310 · Stryker Spine · Aug 18, 2006
- Pioneer Anterior Cervical Plate SystemK043066 · Pioneer Surgical Technology · Feb 3, 2005
Recalls
Recalls FDA ties to K040261, newest first; each opens FDA's recall record| Aug 18, 2008 | Z-0217-2009The Reflex Hybrid 4.0 x 14 mm Variable, Self Tapping Bone Screw, Lot NSZ, was determined to have been anodized with the incorrect color for that screw. | Terminated |
| Jun 14, 2006 | Z-1958-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used. | Terminated |
| Jun 14, 2006 | Z-1960-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used. | Terminated |
| Jun 14, 2006 | Z-1959-2009Design Flaw: The design did not prevent over angulation of the screws when the recommended instrument, All in One Guide, was not used. | Terminated |
Other KWQ Clearances
The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device- corra® cervical plating systemK262149 · Carlsmed, Inc. · Sep 24, 2026
- Palisades Buttress Plate SystemK261309 · Bethesda Medical Devices, LLC · Jul 16, 2026
- ARIES Anterior Cervical Plate SystemsK252776 · ZheJiang Decans Medical Devices Co., Ltd. · May 14, 2026
- CastleLoc-P Anterior Cervical Plate SystemK261112 · L&K BIOMED Co., Ltd. · May 6, 2026
- corra cervical plating systemK260570 · Carlsmed, Inc. · Apr 22, 2026
- VyPlate Anterior Cervical Plate SystemK260697 · Vy Spine, LLC · Mar 27, 2026
- ANTERIS Thoracolumbar Plate SystemK260015 · SpineCraft · Mar 4, 2026
- PathLoc Lumbar Plate SystemK251940 · L&K BIOMED Co., Ltd. · Mar 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Feb 4, 2004
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3060
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files