Arca
Skip to content
Tools
Tools

Roche Diagnostics Urisys 1100 Urine Analyzer

Reagent, Occult Blood. 510(k) K033548, from Roche Diagnostics Corp. (Indianapolis, IN). Product code KHE, class 2.

As of Nov 18, 2026, K033548 is an FDA 510(k) held by Roche Diagnostics Corp., substantially equivalent on Nov 24, 2003 after 12 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 12 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Urisys 1100 Urine Analyzer is a reflectance photometer designed to automatically read and evaluate the results of Chemstrip 5 OB, Chemstrip 7 and Chemstrip 10 MD urine test strips for various urine analytes: specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin and blood.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K033548, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Sep 24, 2018Z-0301-2019Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is…Terminated

Record

Type
Special
Received
Nov 12, 2003
Review panel
Hematology
Regulation
21 CFR 864.6550

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.