Liaison 25 Oh Vitamin D

System, Test, Vitamin D. 510(k) K032844, from DiaSorin, Inc. (Stillwater, MN). Product code MRG, class 2.

As of Nov 21, 2026, K032844 is an FDA 510(k) held by DiaSorin, Inc., substantially equivalent on Feb 12, 2004 after 154 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 154 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The LIAISON$^{®}$ Control 25 OH Vitamin D Set is intended for use as assayed quality control materials to monitor the accuracy and precision of the LIAISON$^{®}$ 25 OH Vitamin D immunoassay.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K032844, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 5, 2008Z-1042-2009DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer.Terminated
Feb 1, 2008Z-1240-2008Inaccurate measurements: Kits may sporadically recover kit and external control values out of range high.Terminated
Feb 14, 2005Z-0638-05An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near…Terminated
Feb 14, 2005Z-0637-05Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator…Terminated

Other MRG clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every MRG device

Record

Type
Traditional
Received
Sep 11, 2003
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1825

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files