Arca
Skip to content
Tools
Tools

Nexiva Closed Iv Catheter System

Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days. 510(k) K032843, from Becton Dickinson Infusion Therapy Systems, Inc. (Sandy, UT). Product code FOZ, class 2.

As of Nov 19, 2026, K032843 is an FDA 510(k) held by Becton Dickinson Infusion Therapy Systems, Inc., substantially equivalent on Feb 27, 2004 after 169 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 169 days
  • Recalls: 17
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Nexiva™ intravascular catheter is inserted into a patient's vascular system for short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port aid in the prevention of needlestick injuries.

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K032843, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 23, 2018Z-0384-2019One lot of the specified needles is not performing as intended due to a failure in the needle tip shield/safety mechanism, which may lead to an increased risk of a contaminated needlestick injury.Terminated
Oct 19, 2009Z-0834-2010, Z-0829-2010, Z-0831-2010, Z-0832-2010, Z-0836-2010, Z-0830-2010, Z-0838-2010, Z-0835-2010, Z-0839-2010, Z-0840-2010, Z-0833-2010, Z-0837-2010Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.Terminated
Feb 3, 2005Z-0585-05, Z-0586-05, Z-0587-05, Z-0584-05The Y-luer connection may crack during normal product use.Terminated

Other FOZ clearances

The newest 510(k)s for the same product code; their median review was 228 days; see every FOZ device

Record

Type
Traditional
Received
Sep 11, 2003
Review panel
General Hospital
Regulation
21 CFR 880.5200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files