Roche Diagnostics Omni S Analyzer
Electrode, Ion Specific, Potassium. 510(k) K032311, from Roche Diagnostics Corp. (Indianapolis, IN). Product code CEM, class 2.
As of Nov 18, 2026, K032311 is an FDA 510(k) held by Roche Diagnostics Corp., substantially equivalent on Oct 17, 2003 after 81 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 17, 2003
- Review time: 81 days
- Recalls: 9
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PO2, PCO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin in samples of whole blood, serum, plasma and aqueous solutions as…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Avl Omni Modular AnalyzerK990092 · Avl Scientific Corp. · Jan 29, 1999
- Omni C AnalyzerK013373 · Roche Diagnostics Corp. · Nov 2, 2001
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc Pack, Roche Combitrol Plus BK111188 · Roche Diagnostics · May 14, 2012
Recalls
Recalls FDA ties to K032311, newest first; each opens FDA's recall record| Feb 22, 2018 | Z-1613-2018The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate. | Terminated |
| Feb 22, 2018 | Z-1612-2018The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate. | Terminated |
| Feb 22, 2018 | Z-1610-2018The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate. | Terminated |
| Feb 22, 2018 | Z-1611-2018The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate. | Terminated |
| Oct 31, 2016 | Z-0673-2017Incorrect Urea results may be generated sporadically on cobas b 221<6>Roche OMNI S6 system and cannot be detected by QC. QC measurement results can be affected as well. This issue is only detectable if comparison result… | Terminated |
| Jun 24, 2015 | Z-2330-2015The affected S2 fluid packs may generate a transponder error when the fluid pack is loaded on to the cobas b 221 system. The error prevents the fluid from being recognized by the system and the instrument cannot start… | Terminated |
| Dec 18, 2014 | Z-1041-2015Chemical component of the F3 Fluid pack and environmental bacterial contamination causes a positive bias up to 39% at 4.1 mmol/l for the glucose measurement and 30% at 4.12 mmol/l for lactate. Urea readings are… | Terminated |
| Sep 12, 2014 | Z-0064-2015The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the… | Terminated |
| Feb 28, 2014 | Z-1315-2014Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected. | Terminated |
Other CEM Clearances
The newest 510(k)s for the same product code; their median review was 273 days; see every CEM device- ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutionsK253491 · Roche Diagnostics · Feb 12, 2026
- ABL90 FLEX PLUS SystemK241037 · Radiometer Medicals Aps · Jan 14, 2025
- EasyStat 300K220396 · Medica Corporation · Sep 26, 2023
- ISE ElectrodesK230890 · Randox Laboratories, Ltd. · Sep 8, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 28, 2003
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files