Arca
Skip to content
Tools
Tools

Duracon Total Knee System

Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer. 510(k) K032163, from Howmedica Osteonics Corp. (Allendale, NJ). Product code MBH, class 2.

As of Nov 18, 2026, K032163 is an FDA 510(k) held by Howmedica Osteonics Corp., substantially equivalent on Sep 12, 2003 after 59 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 59 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Duracon® Total Knee System components are for use in total knee arthroplasty to relieve pain and restore knee function for indications such as: noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular necrosis; rheumatoid arthritis; correction of functional deformity; revision procedures where other treatments or devices have failed; post traumatic…

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K032163, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Dec 1, 2023Z-0916-2024, Z-0918-2024, Z-0917-2024Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.Open

Other MBH clearances

The newest 510(k)s for the same product code; their median review was 153 days; see every MBH device

Record

Type
Traditional
Received
Jul 15, 2003
Review panel
Orthopedic
Regulation
21 CFR 888.3565

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files