Theken Cpod/Lpod Vertebral Body Replacement System

Spinal Vertebral Body Replacement Device. 510(k) K032064, from Theken Surgical,Llc (Akron, OH). Product code MQP, class 2.

As of Nov 17, 2026, K032064 is an FDA 510(k) held by Theken Surgical,Llc, substantially equivalent on Feb 20, 2004 after 232 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Feb 20, 2004
  • Review time: 232 days
  • Recalls: 1
  • Cited as a predicate by: 8 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The CPOD/LPOD is indicated for use in the thoracolumbar spine (i.e. T1 to L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (e.g. fracture). The CPOD/LPOD is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K032064, newest first; each opens FDA's recall record
StartedReasonStatus
Jan 18, 2011Z-2202-2011Certain Theken Spine Vertebral Body Replacement Systems were distributed (i) without certain components (spacers and set screws) that are necessary for their use as vertebral body replacements, and (ii) with a component…Terminated

Other MQP Clearances

The newest 510(k)s for the same product code; their median review was 161 days; see every MQP device

Record

Type
Traditional
Received
Jul 3, 2003
Review panel
Orthopedic
Regulation
21 CFR 888.3060

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.