Oncor Avant-Garde With Coherence Workspaces

Accelerator, Linear, Medical. 510(k) K031764, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.

As of Nov 19, 2026, K031764 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Sep 5, 2003 after 88 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 88 days
  • Recalls: 17
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The intended use of the ONCOR Avant-Garde linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. The ONCOR Avant-Garde includes an Electronic Portal Imaging Device (EPID) that will be marketed as OPTIVUE and is used for the verification of the treatment field and shielding blocks in relation to patient positioning markers and/or anatomical landmarks in…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K031764, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 20, 2019Z-1204-2019Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment…Terminated
Feb 14, 2018Z-1398-2018Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.Terminated
Feb 14, 2017Z-1490-2017Software updateTerminated
Jan 25, 2017Z-1213-2017Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.Terminated
Aug 5, 2015Z-2812-2015A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.Terminated
May 5, 2015Z-0168-2016potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.Terminated
Dec 19, 2014Z-1059-2015A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal…Terminated
Dec 2, 2014Z-0980-2015Use of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in…Terminated
May 28, 2014Z-2174-2014A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to…Terminated
Apr 1, 2014Z-1645-2014Table may lose calibration during patient treatment, which may cause the possibility of mistreatment.Terminated
May 29, 2013Z-1663-2013The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in…Terminated
Nov 4, 2010Z-2642-2011The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.Terminated
Aug 21, 2009Z-1970-2011When editing the note type of a previously entered note, then reverting to the original type, data may be lost.Terminated
Jan 11, 2008Z-1399-2008Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.Terminated
Oct 3, 2007Z-1769-2008Software. The firm issued an Advisory Letter Update instruction TH030/07/S as a result of a complaint that images viewed in the device may shift, which could result in mistreatment. The advisory alerts all its affected…Terminated
Mar 23, 2007Z-0380-2008Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.Terminated
Mar 22, 2007Z-0164-2008Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Jun 9, 2003
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files