Arca
Skip to content
Tools
Tools

Medley ETCO2 Module (Includes Accessories), Model 8300

Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase. 510(k) K031741, from Alaris Medical Systems, Inc. (San Diego, CA). Product code CCK, class 2.

As of Nov 18, 2026, K031741 is an FDA 510(k) held by Alaris Medical Systems, Inc., substantially equivalent on Feb 9, 2004 after 250 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 250 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Medley EtCO _{2 Module is a capnograph that continuously monitors end tidal carbon dioxide (EtCO _{2), fractional inspired carbon dioxide (FiCO _{2), and respiratory rate (RR). The monitor is intended for use by professional healthcare providers where continuous, non-invasive monitoring of these parameters is desired.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K031741, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 1, 2019Z-2443-2019Calibration; it has been determined that the capnographs may have been calibrated during manufacturing using a lower than specified concentration of CO2Terminated
Nov 12, 2015Z-0455-2016CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher…Terminated

Other CCK clearances

The newest 510(k)s for the same product code; their median review was 224 days; see every CCK device

Record

Type
Traditional
Received
Jun 4, 2003
Review panel
Anesthesiology
Regulation
21 CFR 868.1400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.