Arca
Skip to content
Tools
Tools

Endovive Initial Placement Peg Kit, Direct Pej Kit, and Peg Safety Kit

Tubes, Gastrointestinal (And Accessories). 510(k) K031538, from Boston Scientific Corp (Natick,, MA). Product code KNT, class 2.

As of Nov 29, 2026, K031538 is an FDA 510(k) held by Boston Scientific Corp, substantially equivalent on Jun 27, 2003 after 42 days, with 21 recalls tied to it in FDA's recall database.

  • Review time: 42 days
  • Recalls: 21
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K031538, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 20, 2024Z-1517-2024, Z-1516-2024, Z-1515-2024Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a…Open
May 5, 2009Z-1625-2009, Z-1631-2009, Z-1632-2009, Z-1630-2009, Z-1627-2009, Z-1629-2009, Z-1626-2009, Z-1628-2009The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.Terminated
Dec 30, 2008Z-0912-2009, Z-0913-2009, Z-0914-2009, Z-0911-2009The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.Terminated
Jun 2, 2005Z-0966-05, Z-0963-05, Z-0968-05, Z-0965-05, Z-0967-05, Z-0964-05Two blades may be heat staked onto the stiletto instead of the expected one blade.Terminated

Other KNT clearances

The newest 510(k)s for the same product code; their median review was 163 days; see every KNT device

Record

Type
Special
Received
May 16, 2003
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5980

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.