Philips Integris Allura Flat Detector Release 1.2
System, X-Ray, Angiographic. 510(k) K031333, from Philips Medical Systems North America, Inc. (Da Best, NL). Product code IZI, class 2.
As of Nov 20, 2026, K031333 is an FDA 510(k) held by Philips Medical Systems North America, Inc., substantially equivalent on May 13, 2003 after 15 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 25+
- Cited as a predicate by: 2 devices
The Philips Integris Allura Flat Detector release 1.2 system is intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology.
- Philips Integris Allura Flat Detector Release 1.0K022899 · Philips Medical Systems North America Co. · Nov 22, 2002
- Allura Xper FD10K041949 · Philips Medical Systems North America Co. · Jul 30, 2004
- Philips Fresco Release 1K031836 · Philips Medical Systems North America Co. · Aug 14, 2003
| Jan 13, 2025 | Z-1077-2025Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get… | Open |
| Jan 2, 2024 | Z-1151-2024Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure. | Open |
| Feb 19, 2019 | Z-1140-2019, Z-1141-2019Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an… | Terminated |
| Jul 16, 2018 | Z-3219-2018, Z-3215-2018, Z-3216-2018, Z-3220-2018, Z-3217-2018, Z-3218-2018, Z-3214-2018, Z-3213-2018There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension… | Terminated |
| Mar 22, 2017 | Z-1820-2017, Z-1822-2017Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room,… | Terminated |
| Sep 24, 2014 | Z-1280-2016Loss of key image functionality due to a bent pedal of the Footswitch. | Terminated |
| Apr 2, 2012 | Z-1462-2012The monitor Ceiling suspension may fail, which may cause the monitor to drop. | Terminated |
| Jul 8, 2011 | Z-2976-2011System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period. | Terminated |
| Feb 4, 2010 | Z-2449-2010A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit… | Terminated |
| Dec 10, 2008 | Z-0344-2009Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch. | Terminated |
| Oct 23, 2008 | Z-2471-2010A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the… | Terminated |
| Dec 14, 2006 | Z-0370-2007Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested. | Terminated |
| Oct 12, 2006 | Z-0165-2007Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue. | Terminated |
| Feb 6, 2006 | Z-0581-06Potential lock-up of system requiring reboot of system. | Terminated |
| Oct 12, 2005 | Z-0206-06, Z-0207-06Potential for high voltage exposure involving the injector interface connector of patient support tables. | Terminated |
- 780 nm SPY Portable Handheld Imaging (SPY-PHI) SystemK261037 · Stryker Endoscopy · May 28, 2026
- SURGIMAGE SIM 1000H Fluorescence Imaging PlatformK253181 · Surgimage Corporation · May 13, 2026
- IC-Flow™ Imaging System 2.0K243645 · Diagnostic Green GmbH · Sep 22, 2025
- Affirm 800K243077 · Digital Surgery Systems, Inc. (D.B.A True Digital Surgery) · Jun 27, 2025
- Modus IRK231986 · Synaptive Medical, Inc. · Mar 25, 2024
- EXPLORER AIR® II (8001, 8002, 8003); EXPLORER AIR® Sterile Drape (8004)K234090 · Surgvision GmbH · Mar 20, 2024
- Fluorescence Accessories (YELLOW 560 and INFRARED 800 with FLOW 800 Option)K231075 · Carl Zeiss Meditec, Inc. · Jun 20, 2023
- SPY Portable Handheld Imaging (SPY-PHI) SystemK230727 · Novadaq Technologies Ulc (A Part of Stryker) · Jun 5, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Apr 28, 2003
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files