Arca
Skip to content

Navitrack System: Total Hip Replacement Ct-Free Cup, Model 900.204

Neurological Stereotaxic Instrument. 510(k) K030827, from Orthosoft, Inc. (Montreal, CA). Product code HAW, class 2.

As of Nov 18, 2026, K030827 is an FDA 510(k) held by Orthosoft, Inc., substantially equivalent on May 23, 2003 after 70 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: May 23, 2003
  • Review time: 70 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Navitrack™ System – Total Hip Replacement CT-Free Cup is indicated for use as a stereotaxic instrument to assist in the positioning of hip replacement components. It is a computer controlled image-guidance system equipped with a three-dimensional tracking sub-system.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K030827, newest first; each opens FDA's recall record
StartedReasonStatus
Aug 6, 2007Z-2445-2008Inadequate weld on tracker, knob oversized and poor thread design on attachment screw results in seizing. The instrument may break during use, resulting in surgical delay and an increased risk of infection.Terminated

Other HAW Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Traditional
Received
Mar 14, 2003
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.