Alsius Coolgard 3000 & Alsius Catheter System

System, Hypothermia, Intravenous, Cooling. 510(k) K030421, from Alsius Corp. (Irvine, CA). Product code NCX, class 2.

As of Nov 20, 2026, K030421 is an FDA 510(k) held by Alsius Corp., substantially equivalent on Oct 23, 2003 after 255 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 255 days
  • Recalls: 4
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The COOLGARDâ„¢ 3000/Alsius Catheter Thermal Regulation System, using either the Icyâ„¢ or Fortiusâ„¢ model catheter, is indicated for use: - in cardiac surgery patients to achieve and or maintain normothermia during surgery and recovery/intensive care, and, - to induce, maintain and reverse mild hypothermia in neurosurgery patients in surgery and recovery/intensive care. *Page 15* *15* *k030421 2/3*

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K030421, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 9, 2005Z-0369-06Alsius has found that fluid spills or ingress may result in an alarm condition. The potential health risk involved in this state can be the discontinuance of therapy.Terminated
Jul 25, 2005Z-1397-05Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.Terminated
Nov 10, 2004Z-0258-05Product alarms that there may be a power supply interruption.Terminated
Jun 21, 2004Z-1159-04Higher than expected rate of pump alarms occur under specific conditions.Terminated

Record

Type
Traditional
Received
Feb 10, 2003
Review panel
Cardiovascular
Regulation
21 CFR 870.5900

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files