Catheter, Hemodialysis, Non-Implanted. 510(k) K030209, from The Kendall Company (Mansfield, MA). Product code MPB, class 2.

As of Nov 20, 2026, K030209 is an FDA 510(k) held by The Kendall Company, substantially equivalent on May 13, 2003 after 112 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 112 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

MAHURKAR® QPLUS® Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K030209, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 7, 2022Z-0877-2023There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.Open
Apr 29, 2019Z-1955-2019The priming volume values printed on the MAHURKA and Argyle acute catheters and Instructions for Use (IFU) are higher than the volumes required to fill each lumen.Terminated
Dec 2, 2003Z-0234-04Product is mislabeled to catheter length. Product labeled as a 13.5 Fr x 19.5 cm may contain a 13.5 Fr x 16.0 cmTerminated
Dec 2, 2003Z-0233-04Product is mislabeled to catheter length. Product labeled as a 13.5 Fr x 19.5 cm may contain a 13.5Fr x 16.0 cmTerminated

Other MPB clearances

The newest 510(k)s for the same product code; see every MPB device

Record

Type
Traditional
Received
Jan 21, 2003
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5540

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.