Port & Catheter, Implanted, Subcutaneous, Intravascular. 510(k) K030083, from Boston Scientific Corp (Glens Falls, NY). Product code LJT, class 2.
As of Nov 27, 2026, K030083 is an FDA 510(k) held by Boston Scientific Corp, substantially equivalent on Jan 28, 2003 after 19 days, with 21 recalls tied to it in FDA's recall database.
- Review time: 19 days
- Recalls: 21
- Cited as a predicate by: 2 devices
- Nmi PortK122767 · Navilyst Medical, Inc. · Apr 30, 2013
- Vaxcel With Pasv Port (Low Profile), Models 45-236 and 45-238K031844 · Boston Scientific Corp · Jul 15, 2003
| May 13, 2010 | Z-1876-2010, Z-1880-2010, Z-1879-2010, Z-1878-2010, Z-1874-2010, Z-1877-2010, Z-1875-2010FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury. | Terminated |
| Sep 27, 2006 | Z-0631-2007, Z-0630-2007Sterile barrier may have been compromised. Firm received one complaint which demonstrated that the inner and outer seals of the port tray packaging had been compromised. | Terminated |
| Jul 22, 2005 | Z-1474-05, Z-1485-05, Z-1480-05, Z-1477-05, Z-1475-05, Z-1483-05, Z-1473-05, Z-1482-05, Z-1478-05The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use. | Terminated |
| Feb 10, 2004 | Z-0731-04, Z-0732-04, Z-0734-04Reports of catheter separation/fracture after implantation resulting in distal migration of the catheter. | Terminated |
- FLEXI-PORT™ Power Injectable Implantable PortK260916 · Portal Access, Inc. · Jun 17, 2026
- Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable Implantable Port; PowerPort™…K252478 · Bard Access Systems, Inc. · Sep 5, 2025
- PowerPort™ isp Implantable Port; PowerPort™ Slim Implantable Port; PowerPort™ M.R.I. ™ Implantable Port; PowerPort™…K251253 · Bard Access Systems, Inc. · Jun 18, 2025
- PowerPort™ ClearVUE™ Slim Implantable Ports and PowerPort™ ClearVUE™ Slim Implantable PortsK242328 · Bard Access Systems, Inc. · Oct 31, 2024
- PowerPort™ ClearVUE™ Slim ECG Enabled Implantable Port; PowerPort™ ClearVUE™ isp ECG Enabled Implantable Port;…K232737 · Bard Peripheral Vascular, Inc. · Dec 8, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jan 9, 2003
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5965
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files