Arca
Skip to content

Spi Easy Dental Implant Abutment

Abutment, Implant, Dental, Endosseous. 510(k) K023645, from Thommen Medical AG (San Diego, CA). Product code NHA, class 2.

As of Nov 17, 2026, K023645 is an FDA 510(k) held by Thommen Medical AG, substantially equivalent on Nov 13, 2002 after 14 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Nov 13, 2002
  • Review time: 14 days
  • Recalls: 1
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Thommen SPI® EASY Dental Implant Abutments are intended to be used in conjunction with SPI® System dental implants in the maxillary and/or mandibular arch to provide support for crowns, bridges or overdentures.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K023645, newest first; each opens FDA's recall record
StartedReasonStatus
Apr 19, 2007Z-1086-2007The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality…Terminated

Other NHA Clearances

The newest 510(k)s for the same product code; their median review was 169 days; see every NHA device

Record

Type
Special
Received
Oct 30, 2002
Review panel
Dental
Regulation
21 CFR 872.3630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.