Iq 200 System

Counter, Urine Particle. 510(k) K022774, from Intl. Remote Imaging Systems (Chatsworth, CA). Product code LKM, class 2.

As of Nov 21, 2026, K022774 is an FDA 510(k) held by Intl. Remote Imaging Systems, substantially equivalent on Oct 21, 2002 after 61 days, with 14 recalls tied to it in FDA's recall database.

  • Review time: 61 days
  • Recalls: 14
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The iQ 200 System is intended for analysis of urine chemistry, specific gravity, and formed sediment elements, which constitutes typical routine urinalysis. Urinalysis is ordered by physicians as a screening procedure for detection of possible abnormal metabolic or systemic disease, indicated by chemical composition of urine, and of potential renal or urinary tract disease or dysfunction…

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K022774, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 12, 2026Z-0161-2027, Z-0160-2027The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual…Open
Dec 19, 2023Z-0784-2024, Z-0785-2024, Z-0787-2024, Z-0786-2024Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.Open
Apr 5, 2023Z-1597-2023, Z-1599-2023, Z-1598-2023Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.Open
Apr 5, 2023Z-1596-2023Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.Open
Mar 30, 2020Z-1945-2020A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within…Terminated
Oct 28, 2019Z-0914-2020Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification…Terminated
Mar 2, 2018Z-1362-2018Beckman Coulter has determined that there is a potential for under-reporting casts. This can occur if per high-power field (/HPF) units of measurement for casts are selected in the iQ200 software, but the abnormal…Terminated
Nov 10, 2017Z-1366-2018iQ200 Series Urine Microscopy Analyzer may intermittently fail to read urine sample dilution barcode labels causing erroneous results that could delay treatment for health conditions such as hematuria.Terminated

Other LKM clearances

The newest 510(k)s for the same product code; their median review was 164 days; see every LKM device

Record

Type
Traditional
Received
Aug 21, 2002
Review panel
Hematology
Regulation
21 CFR 864.5200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files