Arca
Skip to content

Stryker Navigation System-Knee Module

Neurological Stereotaxic Instrument. 510(k) K022579, from Stryker Leibinger (Kalamazoo, MI). Product code HAW, class 2.

As of Nov 17, 2026, K022579 is an FDA 510(k) held by Stryker Leibinger, substantially equivalent on Jun 2, 2003 after 301 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 2, 2003
  • Review time: 301 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K022579, newest first; each opens FDA's recall record
Jun 5, 2008Z-2288-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2294-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2291-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2289-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2293-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2290-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated
Jun 5, 2008Z-2292-2008Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.Terminated

Other HAW Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Traditional
Received
Aug 5, 2002
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.