Arca
Skip to content
Tools
Tools

The Trabecular Metal Glenoid-Bigliani/Flatow The Complete Shoulder Solution, Model 4306

Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented. 510(k) K022377, from Implex Corp. (Allendale, NJ). Product code KWT, class 2.

As of Nov 18, 2026, K022377 is an FDA 510(k) held by Implex Corp., substantially equivalent on Dec 10, 2002 after 141 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 141 days
  • Recalls: 8
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; un-united humeral head fractures of long duration; irreducible 3- and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K022377, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 23, 2005Z-0024-06, Z-0023-06, Z-0028-06, Z-0027-06, Z-0029-06, Z-0025-06, Z-0026-06, Z-0030-06The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.Terminated

Other KWT clearances

The newest 510(k)s for the same product code; their median review was 232 days; see every KWT device

Record

Type
Traditional
Received
Jul 22, 2002
Review panel
Orthopedic
Regulation
21 CFR 888.3650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.